MULTI PARAMETER PROBE - P230 SMP Description: The multi parameter probe SMP allow the simultaneous and continuous measuring of the most common surface or groundwater physical–chemical characteristics.
Electrodes: The electrodes are exchangeable and are selected from the high accuracy, long-term stable sensors available in the market. The probe can include the following sensors: pH-value, electric conductivity, redox-potential, dissolved oxygen, temperature and water level.
Internal data-logger: The probe electronic management is entrusted to an internal data logger that achieve among other the following functions: signal acquisition and A/D conversion, data storage and/or data transmission to a remote station located at the ground or at some distance. The probe configuration and all functionalities can be modified locally or remotely programmed using the user friendly Olimposw software package.
Data downloading: Digital output for local connection to external data logger, Laptop or remote data transmission using any standard data transmission device. The ASCII format allow to import the data in Excel or any other standard package.
Operation: The probe can be submerged up to water column of 30.0 m, it’s connected to the surface control unit using a self-supporting cable for power supply and data transmission. Optionally special transducer cable with barometric pressure vent
Housing: The probe body is realized in special AISI 316 stainless steel that warranty high mechanical robustness and high stability on aggressive environmental conditions. |
Technical data Temperature
sensor: Pt1000 thermo resistance DIN standard
Range: -5 ÷ 60 °C
Accuracy: ±0,2 °C
Resolution: 0,1 °C
PH-value
Sensor: glass electrode with internal Ag/AgCl reference
Measure technique: potentiometric based on combined electrode with independent reference electrode
Standard range: 2 ÷ 12 units
Accuracy: ±0,1 units
Resolution: 0,05 units
Operational temperature: -5 ÷ 60 °C
Compensation: automatic for temperature 0 ÷ 40 °C
Calibration method: with reference solution and internal sw
Redox-potential:
Sensor: glass electrode with internal Ag/AgCl reference
Measure technique: potentiometric
Standard range: -1000 ÷ 1000 mV
Accuracy: ± 25 mV
Resolution: 5 mV
Calibration method: with reference solution and internal sw
Conductivity
Measure technique: Conductimetric with 3 black platinum electrodes cell
Standard range: 0÷3000 µS/cm ; 0÷30000 µS/cm (autoranging)
Accuracy: ± 30 µS ; ± 300 µS
Resolution: 2 µS ; 20 µS
Operational temperature: -5 ÷ 60 °C
Compensation: automatic for temperature: 0 ÷ 40 °C
Calibration method: with reference solution and internal sw |  |